ACR Electronics SmartReader Plus Bedienungsanleitung Seite 78

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 103
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 77
SmartReader Data Loggers
70
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
4. Cover the bucket with foam chips or another suitable insulating material. Let sit for at
least five minutes while the temperature stabilizes and the thermistor has a chance to
respond. If the test is to continue for more than a few minutes, add more ice periodically,
as before, insuring that it is tightly packed to the bottom of the bucket each time. The
goal is to ensure that the thermistors are always in contact with an ice / water mixture
over their entire surface.
5. Measure the temperature of the water (it should be very close to 0°C) and the resistance
of the thermistor. This is the Low calibration point data.
6. Next, fill up a pot of water and heat it up till it reaches approximately 50°C. Once the
temperature and thermistor have stabilized, again take an accurate reading of both the
temperature and the thermistor's resistance, and record. This is the Mid calibration point
data,
7. Now heat the water so it boils. Record the temperature (100°C) and the resistance again.
This is the Mid calibration point data.
8. With TrendReader software, create a thermistor equation (using the equation wizard)
from the Equation Menu. Save the equation to the logger.
Note that the accuracy of the above test procedure is primarily limited by the cumulative
accuracy of the instruments you are using and how carefully you carry out the procedure. You
should, however, be able to achieve accuracies within ±1.0°C under most circumstances for
thermistors with R25 values of between 10k and 100k ohms. You can perform tests with
thermistors of other R25 values by using dry ice (for very low temperatures down to -70°C) or
an oven (for temperatures up to a maximum of 450°C).
To avoid damaging your thermistors, when calibrating, make sure to note the following:
· do not exceed the upper temperature limit rating for your high calibration point, and
· do not introduce the sensors to rapid changes in temperature (for example, from the ice
bucket right into the boiling water).
1.14.3 Relative Humidity Calibration
If you want to confirm the calibration of your RH sensor or EH-020A Temperature and Relative
Humidity probe, you can simply return it to your authorized sales representative or perform
your own field test using the special adjustment provisions in TrendReader software.
Recommended methods of calibration include using an accurate relative humidity test
chamber or by mixing saturated salt solutions. Both tests are performed at an ambient room
temperature of 20 to 25°C (68 to 77°F).
RH Chamber Test
If you intend to use a Relative Humidity Test Chamber for your calibration, make sure it has
been recently tested to within ±1%. The procedure to calibrate is as follows:
1. Set up your humidity data logger with a sample rate of 20 seconds and equation 38 on
the RH channels. Make sure you have selected Normalized Byte data units. If you are
Seitenansicht 77
1 2 ... 73 74 75 76 77 78 79 80 81 82 83 ... 102 103

Kommentare zu diesen Handbüchern

Keine Kommentare