ACR Electronics SmartReader Plus Bedienungsanleitung Seite 59

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 103
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 58
SmartReader Data Logger
51
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
current channels have a low 100 ohm impedance). This helps to minimize the voltage
requirements and power consumption of the circuit. Check the specifications for the
transducers you intend to use to ensure the SmartReader 7's input impedances are
acceptable.
Two-Wire Transmitters
Two-wire transmitters are transducers that form part of a current loop circuit. They vary the
current flow in accordance with changes in the variable which they are sensing. Most two-wire
transmitters have a 4-20mA output.
Three-Wire Transducers
Three-wire transducers have three connections: a supply (excitation) voltage input (V+IN), an
output voltage (V+OUT) and a common (COM-). You can usually tell if a transducer is 3-wire
by reviewing the wiring diagram and looking for a single common connection.
Four-Wire Transducers
Four-wire transducers fall into two categories: line-type and Wheatstone bridge types. No
more than one line-type transducer can be attached to a SmartReader 7 unless it has an
isolated output (no electrical connection to the line). Bridge-type transducers can only be
hooked to a SmartReader 7 if they are signal conditioned.
Output
The specified output of a transducer or transmitter will determine whether it is compatible for
use with your SmartReader 7. Some of the most popular transducer outputs (such as 0 to
2.5V or 0 to 5V DC) are directly compatible with the logger. The most popular transmitter
output is 4 to 20 mA which can be easily used with the SmartReader 7.
1.9.7 Specifications
Voltage and Current Inputs
Standard Ranges: 0 to 2.5 volts DC (1 channel)
0 to 5 volts DC (2 channels)
0 to 10 volts DC (1 channel)
0 to 200mV DC (1 channel)
0 to 25mA (2 channels)
Accuracy: ±1% F.S.
Input Impedance: >1 M ohms (0 to 200 mV channel)
>1 M ohms (0 to 2.5 V channel)
20 k ohms (0 to 5.0 V channels)
40.9 k ohms (0 to 10.0 V channel)
100 ohms (0 to 25 mA channels)
Maximum Input Voltage: Voltage channels: ±40 V
(reverse polarity protected)
Current channels: ±70 mA
(reverse polarity protected)