1.5.2 Setup and Use
The basic procedure for setting up your SmartReader 3 is as follows:
1. Connect one or more current probes, specified above, to the current channel inputs on
your SmartReader 3. Make sure to observe polarity when making connections. Connect
the current probe's red (+) wire to the channel's "+ Red" input. Connect the current
probe's black wire to the channel's "- Black" input.
Figure 3-2: Connecting an external current probe
2. Adjust the range setting on the current probe(s) to suit your application. A good way to
verify a conductor's maximum range is to check the circuit breaker or fuse rating. Set the
range on your current probe to suit. Since the probes do not give predictable readings
below 10%, always choose the lowest range possible that will encompass the full range
of data you want to record.
3. Make sure to enable each channel you want to use. If you are not using a channel,
disable it to save logger memory.
Channel
Description
Equation
0
Internal Temperature
45
1
Current 1
as below
2
Current 2
as below
3
Current 3
as below
A60FL / A65FL
Switch
Range
Equation
5
0.5 to 5 A
63